01 / 12
Product introduction
Features
Technical characteristics:
●Automatic leak detection: inflate the cabin and check whether the system is closed. If the airtight requirements are met, the sterilization process will be
started. If the airtight requirements are not met, continue to inflate and seal, and then leak detection, and so on, until the airtight conditions are met.
● Sterilization liquid shortage alarm: automatically judge whether the sterilization liquid is sufficient for a complete sterilization process. If the sterilization
liquid is in sufficient, an alarm will be prompted.
●Unique dehumidification method:dehumidifier is not required to maintain low humidity in the cabin during preheating and sterilization.
●Efficient flash method: high-precision flash temperature control enables hydrogen peroxide to decompose and evaporate efficiently during flash.
● Rapid decomposition and residue removal: the efficient hydrogen peroxide plasma decomposer can not only decompose the hydrogen peroxide
suspended in space, but also decompose the hydrogen peroxide on the surface. Except for the residue, it is fast and complete, and does not need to be
discharged.
●Record the whole process of sterilization: automatically detect the sterilization concentration and automatically draw the sterilization process curve.
● Double-door electronic interlock function: when the equipment is in the state of leak detection, sterilization and residue removal, both doors cannot be
opened. Only after the residue removal reaches the safe concentration can the side door of high cleanliness level be opened. The two doors cannot be
opened at the same time.
●Bacteriostatic treatment of the interior surface of the cabin:the interior surface is treated with special treatment, which can achieve sterilization and
bacteriostasis, and make the interior surface free of microbial breeding ground forever.
●Software system conforming to computerized verification: the system has the function of audit trail and electronic signature. Comply with the audit
requirements of FDA 21CFR Part 11.
·Configurable ultraviolet disinfection lamp:ultraviolet disinfection is used when VHP sterilization is not required.
Model description

Model specifications
Inner cabin dimensions Exhaust air interface fresh air inlet power supply Maximum Power
Specifications/Model Overall dimensions(W*D*H)mm
Consumption
(W*D*H)mm
VHPB-400/500/500
VHPB-500/500/500
VHPB-600/600/600
VHPB-700/700/700
VHPB-800/800/800
VHPB-900/900/1000
1600*1100*2250 1000*1000*1000 φ133 /50HZ
VHPB-1000/1000/1000
φ133 /50HZ
VHPB-1200/1000/1000 1800*1100*2250 1200*1000*1000
VHPB-1200/2000/1800 1800/2100/2450
Configurations and specifications

Downloads
No downloads are available yet.
Request product information