Models and specifications

STI-1200

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

1600*880*2250

Operation cabin Size(W*D*H)mm

1195*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1200/BPB600

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

1600*880*2250

Operation cabin Size(W*D*H)mm

1195*680*700

Size of transfer cabin(w*D*H)mm

600*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1500D/BPB600

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

2100*1200*2250

Operation cabin Size(W*D*H)mm

1195*680*700

Size of transfer cabin(w*D*H)mm

600*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1800

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

2200*880*2250

Operation cabin Size(W*D*H)mm

1795*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1800/BPB600

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

2200*880*2250

Operation cabin Size(W*D*H)mm

1795*680*700

Size of transfer cabin(W*D*H)mm

600*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1800D/BPB600

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

2200*1200*2250

Operation cabin Size(W*D*H)mm

1795*1195*700

Size of transfer cabin(W*D*H)mm

600*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM

STI-1800/PB600

Product Introduction

The sterile isolator is a closed-type operating device designed to provide a localized Class A clean environment, primarily used for aseptic operations, microbial control, and the protection of operators or products from contamination. It is widely applied in pharmaceuticals, biotechnology, healthcare, and laboratory settings—particularly in scenarios such as sterile pharmaceutical production, cell culture, and manipulation of highly pathogenic microorganisms. The isolator employs hydrogen peroxide vapor (VHP) sterilization technology to comprehensively sterilize the chamber, achieving a kill rate of up to 6-log. By integrating a HEPA/ULPA filtration system, a sealed chamber design, and glove-based operations, the device ensures that the operating environment complies with regulatory requirements, including GMP and ISO 14644-1 standards.

Technical parameter

project

technical parameter

Overall dimension(W*D*H)mm

1600*880*2250

Operation cabin Size(W*D*H)mm

1195*680*700

Size of transfer cabin(W*D*H)mm

600*680*700

Compressed air volume

≥0.6Mpa

Work area cleanliness

Class A

Air supply velocity

0.36-0.54 m/s

noise

≤75dB(A)

illuminance

>300Lux

power

4KW

Power Supply Voltage

AC220V±10%,50Hz±1Hz

control model

Manual mode, Automatic mode

Door sealing method

Inflatable sealing

sterilization method

Vaporized hydrogen peroxide

Sterilization efficiency

6lg

Automatic sterilization duration

<240 min

Concentration of hydrogen peroxide after residual removal

<1PPM